Specifications

*TD-SC1 (485) only
**TD-SC1 (D/A) only
**TD-SC1 (CCL) only
***TD-SC1 (E/IP) only

  • Specifications and appearance are subject to change without notice.
  • Weight and dimensions are approximate.
  • Illustrations in this owner's manual might differ slightly from production models.
Bridge Voltage DC 10 V, 5 V±10% (30mA current maximum, can be used with remote sense)
Signal input range ±5.0mV/V
Equivalent Input/TEDS Calibration range 0.05mV/V - 5.0mV/V
Calibration precision Within 0.1% F.S. (when using 1m standard TEAC Ø8, 6-core shielded cable with 350Ω impedance, 10V BV and 5mV/V setting)
Precision Linearity Within 0.01%F.S. +1 digit (when input is 5mV/V)
Zero drift Within 0.5µV/ºC (Input conversion value)
Gain drift Within ±0.005%F.S/ºC
A/D conversion 24-bit, 20000 times/second
Digital filter Select 3, 10, 30, 100, 300, 1000 Hz(−6 dB/oct) or none
D/A output** Output with same frequency as A/D conversion, isolated output, ±1–±10V output (set in 1V steps) and about 1/59000 resolution (when set to ±10 V), or 4–20mA current output and about 1/43000 resolution
TEDS function IEEE1451.4 class 2 mix mode interface
Indicator value Display range −99999 to 99999
Decimal point Position selectable
Times displayed Select 4, 6, 10 or 20 times/second
Setting items Calibration settings Zero calibration/span calibration (TEDS calibration, actual load calibration, equivalent input calibration)
Function settings High limit, low limit, comparison mode, hysteresis, nearly zero, moving average, low pass filter, motion detect, zero tracking, static strain, digital zero, digital offset, zone definition, hold mode, lock, minimum grid, display times, bridge voltage, digital zero limit, clear digital zero, comparison output pattern, select data output**, D/A converter**, remote sense
Hold functions Sample hold, peak hold, bottom hold, peak to peak hold, peak and bottom hold, average hold, zone definition hold (peak, bottom, peak to peak, peak and bottom, average)
External input and output signals Input Hold, clear, digital zero (isolated from main unit circuits using a photocoupler)
Output HI, OK, LO open collector output (isolated from main unit circuits using a photocoupler)
RS-485* A+, B–(isolated from main unit circuits using a photocoupler), TRM, SG
CC-Link** DA, DB (isolated from main unit circuits using a photocoupler), DG, SLD
EtherNet/IP*** P1, P2
Power supply Ratings: 24V DC ±10% 7W
Operating temperature range 0℃ to 40℃
Storage temperature range –20℃ to 60℃
Operating humidity range 85% RH or less (without condensation)
Applicable standards CE marking, FCC (Class A), UL61010-1, UKCA marking
External dimensions (W×H×D) Approximately 30 mm × 85 mm × 110 mm (without protrusions)
Weight About 210 g

Included accessories

  • Input/output connector plug B2CF 3.50/18/180LR SN OR BX
    1
  • Micro USB-B cable
    1
  • Power input terminal bank cover (pre-installed on unit)
    1
  • The following additional accessories are included with the TD-SC1 (D/A).
    KEC-NS0604-02 D/A terminal bank
    1
  • D/A terminal bank cover
    1
  • The following additional accessories are included with the TD-SC1 (485).
    KEC-NS0604-02 RS-485 terminal bank
    1
  • RS-485 terminal bank cover
    1
  • The following additional accessories are included with the TD-SC1 (CCL).
    KEC-NS0604-02 CC-Link terminal bank
    1
  • CC-Link terminal bank cover
    1

Options

AC adapter (For use in Japan only, AC100V-240V)
AC adapter (For use in Japan only, AC100V-240V)