LOAD ADC |
ADC plus over |
− LOAD ADC |
ADC minus over |
FULL |
Display plus over (greater than maximum display value) |
− FULL |
Display minus over (less than minimum display value) |
OVER FULL |
Input is exceeding maximum input range (5.2 mV/V) |
− OVER FULL |
Input is less than minimum input range (−5.2 mV/V) |
ZERO OVER |
Zero balancing range exceeds regulated values |
ZERO ERROR |
Zero-adjusting failed within specified time |
OUTPUT CAL OVER |
Sensor output exceeds calibration range |
OUTPUT CAL SHORT |
Sensor output does not achieve calibration range |
MINUS INPUT |
Sensor input is negative |
TEDS READ ERROR |
A valid TEDS sensor is not connected |
PARAMETER ERROR |
Irregular setting value exists |
R.O.SET OVER |
Rated output exceeds the set range (greater than 5.0 mV/V) |
R.O.SET SHORT |
Rated output below the set range (less than 0.3 mV/V) |
ZEROLIMIT OVER |
Digital Zero Limit exceeded |
ERROR |
An error has occurred |
DA OVER |
DA output is outside output range |
DA - OVER |
DA output is outside output range |
SYSTEM ERROR |
A system error has occurred |
INVALID OPERATION |
Operation is invalid |
MEMORY FULL |
Memory for saving data is full. Delete data or enable overwriting using "System settings 2" |