Indicator spec list of all models
Specifications |
Color Graphics Digital Indicator
TD-9000T
|
Digital Indicator
TD-700T
|
Digital Indicator
TD-260T
|
Signal Conditioner
TD-SC1
|
Portable Digital Indicator
TD-01Portable
|
---|---|---|---|---|---|
Communication interface | EtherNet/IP,CC-Link,RS-232C (D-sub 9-pin), USB | CC-Link,RS-485 | RS-232C | RS-485,EtherNet/IP,CC-Link | |
Global support (safety standard) | CE marking, FCC (Class A), UL61010-1, UKCA | CE marking EN61326 (class A), UL61010-1 | Installation class I, EN61010-1 Pollution degree 2 Overvoltage category II, CE marking | CE marking, FCC (Class A), UL61010-1, UKCA | CE marking, VCCI (Class A), FCC (Class A) |
Language | Japanese / English / Chinese / Korean | Japanese / English | × | × | Japanese / English |
Display | 4.3" color LCD (480×272) | 320 x 240 color liquid crystal | 14.6 mm high alphanumeric, shown by 7-segment green LED display | × | 2.4" color TFT LCD |
A/D conversion | 24-bit, 5000 times/second, 25000 times/second | 24-bit, 4000 times per second, 20000 times per second (fast sampling mode) | 16-bit,100 times/sec | 24-bit, 20000 times/second | 1000 times/second, 24-bit |
Screen | Digital load value, Waveform, Archive data, Setting | Waveform & Bar meter display function,Digital load value | Digital load value | × | Setting screens, indicator value digital display, graph display, recorded data list display, static strain display |
Display range | −32000 to 32000 | −99999 – 99999 | −19999 to 99999 | × | -99999 to 99999 |
Number of input channels | 2 | 1 | 1 | 1 | 1 |
Signal input range | ±3.2 mV/V | ±3.2 mV/V | ±3.2 mV/V | ±5.0mV/V | ±5.0mV/V |
Bridge voltage | DC, 2.5/5/10 V, ±10% (30mA current maximum, can be used with remote sense) | DC, 2.5/10 V, ±10% (30mA current maximum, can be used with remote sense) | DC, 2.5/10V, ±10% (Current: 60mA max. Remote sensing available) | DC 10 V, 5 V±10% (30mA current maximum, can be used with remote sense) | DC, 2.5 V ±5% (30mA maximum current) |
Calibration precision | Within 0.1% F.S. (when using 1m standard TEAC Ø8, 4-core shielded cable with 350Ω impedance, 10V BV and 3.0mV/V setting) | Within 0.1% F.S. (when using 1m standard TEAC Ø8, 4-core shielded cable with 350Ω impedance, 10V BV and 3.0mV/V setting) | Within 0.1% FS (at 10V for BV and 0.5mV/V or more for sensor sensitivity) | Within 0.1% F.S. (when using 1m standard TEAC Ø8, 6-core shielded cable with 350Ω impedance, 10V BV and 3.0mV/V setting) | Within 0.1% F.S. (when using a 1m standard TEAC Φ8,6-core shielded cable with 350Ω impedance, when 5mV/V) |
Linearity | Within 0.01% F.S. +1 digit (when input is 3.0 mV/V) | Within 0.01% F.S. +1 digit (when input is 1 mV/V or greater) | Within 0.01% FS + and 1 digit (at an input of 1 mV/V or more) | Within 0.01% F.S. +1 digit (when input is 5.0 mV/V) | Within 0.01% F.S. + 1 digit (when 5mV/V) |
TEDS function | IEEE1451.4 class 2 mix mode interface | IEEE1451.4 class 2 mix mode interface | IEEE1451.4 class 2 mix mode interface | IEEE1451.4 class 2 mix mode interface | IEEE1451.4 class 2 mix mode interface |
Static strain display | 〇 | 〇 | 〇 | 〇 | 〇 |
Disconnection detection | 〇 | 〇 | × | 〇 | 〇 |
Low-pass filter | Select 3, 10, 30, 100, 300, 1000 Hz(−6 dB/oct) or Off | Select 3 Hz (−6dB/oct), 10, 30, 100, 300, 1000 Hz (−12dB/oct) or none | Selectable from 3, 10, 30, 100, 300 Hz (−12dB/oct) | Select 3, 10, 30, 100, 300, 1000 Hz(−6 dB/oct) or none | Moving average (select from OFF, 16, 32, 64, 128, 256, 512, 1024, 2048) |
D/A output | Output with same frequency as A/D conversion, isolated output, ±1–±10V output (set in 1V steps) and about 1/59000 resolution (when set to ±10 V), or 4–20mA current output and about 1/43000 resolution | 4000 times per second, isolated output, ±1–±10V voltage output (set in 1V steps) and about 1/59000 resolution (when set to ±10V), or 4–20mA current output and about 1/43000 resolution | 0 - ±10V or 4 - 20mA | 4000 times per second, isolated output, ±1–±10V voltage output (set in 1V steps) and about 1/59000 resolution (when set to ±10 V), or 4–20mA current output and about 1/43000 resolution | ±2.0V |
Displacement(Pulse) | Pulse type:A/B phase or A phase, differential square wave (RS-422 conformance) Maximum input frequency:2MHz Maximum count number:1500000 Power output:+5V ±10%, 500mA |
× | × | × | × |
Displacement(Voltage) | Input:±5.2 V Low-pass filter:10/30/100/300Hz Power output:DC 12V ±10%, 250 mA |
× | × | × | × |
Real-time judgments | 〇 | × | × | × | |
Band judgment | Offset reference band / Designated value band | × | × | × | × |
Hold setting | Constant comparison, sampling, peak, bottom, peak to peak, maximum/minimum, inflection point and average value | Sample hold, peak hold, bottom hold, peak to peak hold, peak and bottom hold, average hold, zone definition hold (peak, bottom, peak to peak, peak and bottom, average) | Sample hold, Peak hold, Bottom hold, Peak-to-Peak hold, block-specified hold (Peak, Bottom, Peak-to-Peak) | Sample hold, peak hold, bottom hold, zone definition hold (peak, bottom) | Sample hold, peak hold, bottom hold, zone definition hold (peak, bottom) |
External output signals | Load: HH / HI / OK / LO / LL Displacement: HI / OK / LO |
HH, HI, OK, LO, LL open collector output (isolated from main unit circuits using a photocoupler) | 1a contact for each relay: HH, HI, OK, LO, LL, analog voltage | HI, OK, LO open collector output (isolated from main unit circuits using a photocoupler) | × |
Data Record | 〇 | × | × | × | △(Built-in memory) |
Recording media | SD/ SDHC(2~32GB, Class 10 recommended) | × | × | × | × |
Operating temperature range | 0°C to 40°C | 0°C to 40°C | -10°C to 40°C | 0°C to 40°C | 0°C to 40°C |
External dimensions (W × H × D) | Approximately 114 mm × 96 mm × 140 mm (without protrusions) | Approximately 96 mm × 53 mm × 132 mm (without protrusions) | Approx. 96 mm × 96 mm × 146 mm (protrusions not included) | Approximately 30 mm × 85 mm × 110 mm (without protrusions) | Approximately 85mm × 140mm × 35mm (without protrusion) |
Weight | About 960g | About 300g | Approx. 950g | About 210g | About 320g (including batteries) |
Power supply | Ratings: 24V DC ±10%, 13 W | Ratings: DC 12–24 V, 9 W | 100 to 240V ±10% 50 - 60Hz AC, 13W | Ratings: 24V DC ±10% 7W | 4 alkaline or NiMH AA batteries, USB bus power (built-in Micro-USB B connector) |