Load |
Bridge voltage |
DC, 2.5/5/10V, ±10% (30mA current maximum, can be used with remote sense) |
Signal input range |
±3.2mV/V |
Equivalent input/TEDS |
Calibration range |
0.1mV/V - 3.2mV/V |
Calibration precision |
Within 0.1% F.S. (when using 1m standard TEAC Ø8, 6-core shielded cable with 350Ω impedance, 10V BV and 3.0mV/V setting) |
Precision |
Linearity |
Within 0.01% F.S. +1 digit (when input is 3.0mV/V) |
Zero drift |
Within 0.5µV/°C (Input conversion value) |
Gain drift |
Within ±0.005% F.S./°C |
A/D conversion |
24-bit, 5000 times/second, 25000 times/second |
Low-pass filter |
Select 3, 10, 30, 100, 300, 1000 Hz(−6 dB/oct) or Off |
D/A output |
Output with same frequency as A/D conversion, isolated output, ±1–±10V output (set in 1V steps) and about 1/59000 resolution (when set to ±10 V), or 4–20mA current output and about 1/43000 resolution |
TEDS function |
IEEE1451.4 class 2 mix mode interface |
Displacement |
Pulse |
Pulse type |
A/B phase or A phase, differential square wave (RS-422 conformance) |
Maximum input frequency |
2 MHz |
Maximum count number |
15000000 |
Power output |
+5V ±10%, 500mA |
Voltage |
Input |
±5.2V |
Low-pass filter |
10/30/100/300Hz |
Power output |
DC 12V ±10%, 250mA |
Display |
4.3" color LCD (480×272) |
Indicator value |
Display range |
−32000 to 32000 |
Decimal point |
Display position selectable |
Times displayed |
4 times/second |
External input and output signals |
Input |
Differential pulse position sensor (A phase, B phase), force backlight lighting, prevent touchscreen operation, force reset, work switching (4-bit), switch zone, clearresults (reset measurement results), enable/disable judgment output, start/stop measurement, zero balance displacement, digital zero Isolated from main unit circuits using a photocoupler |
Output |
Load judgment output (HH, HI, OK, LO, LL), displacement judgment output (HI, OK, LO), load cell error, measurement complete, trigger output (1, 2) Open collector output (isolated from main unit circuits using a photocoupler) |
RS-232C |
RXD, TXD |
Power supply |
Ratings: 24V DC ±10%, 13W |
Operating temperature range |
0°C to 40°C |
Storage temperature range |
−20°C to 60°C |
Operating humidity range |
85% RH or less (without condensation) |
Applicable standards |
CE marking, FCC (Class A), UL61010-1, UKCA marking |
External dimensions (W × H × D) |
Approximately 114mm × 96mm × 140mm (without protrusions) |
Weight |
About 960g |