TEAC Load Cell site

Specifications

*Specifications and appearance are subject to change without notice.
*Weight and dimensions are approximate.
*EtherNet/IP・CC-Link and USB/RS232C can be used for transmission with this unit. When using EtherNet/IP・CC-Link for control, however, discontinue control by USB/RS232C, and when using USB/RS232C for control, discontinue control by EtherNet/IP・CC-Link. Operation cannot be guaranteed when controlled by both at the same time.

Load Bridge voltage DC, 2.5/5/10V, ±10% (30mA current maximum, can be used with remote sense)
Signal input range ±3.2mV/V
Equivalent input/TEDS Calibration range 0.1mV/V - 3.2mV/V
Calibration precision Within 0.1% F.S. (when using 1m standard TEAC Ø8, 6-core shielded cable with 350Ω impedance, 10V BV and 3.0mV/V setting)
Precision Linearity Within 0.01% F.S. +1 digit (when input is 3.0mV/V)
Zero drift Within 0.5µV/°C (Input conversion value)
Gain drift Within ±0.005% F.S./°C
A/D conversion 24-bit, 5000 times/second, 25000 times/second
Low-pass filter Select 3, 10, 30, 100, 300, 1000 Hz(−6 dB/oct) or Off
D/A output Output with same frequency as A/D conversion, isolated output, ±1–±10V output (set in 1V steps) and about 1/59000 resolution (when set to ±10 V), or 4–20mA current output and about 1/43000 resolution
TEDS function IEEE1451.4 class 2 mix mode interface
Displacement Pulse Pulse type A/B phase or A phase, differential square wave (RS-422 conformance)
Maximum input frequency 2 MHz
Maximum count number 15000000
Power output +5V ±10%, 500mA
Voltage Input ±5.2V
Low-pass filter 10/30/100/300Hz
Power output DC 12V ±10%, 250mA
Display 4.3" color LCD (480×272)
Indicator value Display range −32000 to 32000
Decimal point Display position selectable
Times displayed 4 times/second
External input and output signals Input Differential pulse position sensor (A phase, B phase), force backlight lighting, prevent touchscreen operation, force reset, work switching (4-bit), switch zone, clearresults (reset measurement results), enable/disable judgment output, start/stop measurement, zero balance displacement, digital zero Isolated from main unit circuits using a photocoupler
Output Load judgment output (HH, HI, OK, LO, LL), displacement judgment output (HI, OK, LO), load cell error, measurement complete, trigger output (1, 2) Open collector output (isolated from main unit circuits using a photocoupler)
RS-232C RXD, TXD
Power supply Ratings: 24V DC ±10%, 13W
Operating temperature range 0°C to 40°C
Storage temperature range −20°C to 60°C
Operating humidity range 85% RH or less (without condensation)
Applicable standards CE marking, FCC (Class A), UL61010-1, UKCA marking
External dimensions (W × H × D) Approximately 114mm × 96mm × 140mm (without protrusions)
Weight About 960g

Included accessories

 
  • SENSOR connector plug
    B2CF 3.50/18/180LR SN OR BX or equivalent part
    1
  • CONTROL connector plug
    HDCB-37P (05) connector or equivalent part
    HDC-CTH (4-40) (10) case or equivalent part
    Power input terminal bank cover(pre-installed on unit)
    1
    1
    1